By adopting the versatile soil corrosion rate measurer, the measurement of soil parameters is achieved by a versatile sensor, i.e. a probe.



Technique Features:

·A combined electrochemical probe assembled with 5 electrode collars of the same material is adopted and effectively reduces the effect of soil IR drop when measuring the corrosion current density; the connection method of counter electrode (CE), working electrode (WE) and reference electrode (RE) is represented by (CE-WE-RE), and the connection method of combined electrochemical probe is represented by (1-5,2-4,3). The area of working electrode (WE) and reference electrode (RE) is 5cm2;

·With four-electrode probe and a relatively low measuring voltage, the soil resistivity can be accurately measured, and the readout difficulty brought by high voltage measurement and physicochemical performance damage caused by soil electrolysis can be avoided;

·Standard platinum electrode and copper sulphate electrode are used to measure the soil oxidation-reduction potential;

·High-precision temperature measurer is fixed on the PTFE insulating rod of probe to measure soil temperature;

·Preformed hole method does little damage to the original soil structure and no excavation is required;

·Tapered design of front end is adopted to ensure good contact between probe and soil;

·Probe can be reused by polishing.


Equipment Application Fields

· Soil corrosion evaluation
· External corrosion evaluation of buried pipelines for oil and gas delivery and urban buried pipelines
· Concrete corrosion evaluation
· Corrosion evaluation of corrosion medium in aqueous solution
· and so on.

Technical Index:
·Ambient conditions of instrument during normal operation: Temperature: 0℃ to 40℃
Relative humidity: ≤80%RH
·Measurement range of soil temperature: -10℃ to 100℃, error: 2℃
·Measurement range of soil oxidation-reduction potential: ±2000mV, error: ≤1%
·Measurement range of soil resistivity: 0 to 10KΩm,error:≤2%
·Measurement range of soil polarization resistance: 40Ω to 300KΩ,error:≤3%
·Measurement range of soil corrosion rate: 1×10-4 to 2 mm/a
·The charging time of storage battery is 6 hours, during which it can function properly. The battery can continuously work for 24 hours after one full charge.

Configuration Information:
NO. Name Type Qty.
1 Main unit of measurer ZKW180 1
2 Versatile soil probe ZKW120/Corrosive material 1
3 Oxidation-reduction electrode ZKW121 1
4 Copper sulphate electrode GY250 1
5 Preformed hole machine ZKW122 1
6 Soil resistance and corrosion rate checker ZKW130/10K 1
7 Oxidation-reduction potential checker ZKW131/1V 1
8 Charger (12V-4A) GY450 1
9 Connecting line of versatile soil probe ZKW150 1
10 Connecting line of oxidation-reduction probe ZKW151 1
11 Communication line (USB dual male ) GY560 1
12 Instruction manual ZKW190 1
13 CD ZKW191 1

Packing Description:

Net weight of the main unit:
Gross weight of the main unit:
Packing dimensions:
Wrap: Cardboard boxes for outer packing and aluminum boxes for inner packing.